LB4200 Multi-Detector Low Background Alpha/Beta Counting System
It can collect water samples, biological samples, environmental samples and other samples α、 total β、 Activity measurement of Cs-137 and Pb-210. It is widely used in disease control and Prevention Center, water supply company, product quality supervision and inspection unit, environmental monitoring station, advanced laboratory and other units.
Canberra lb4200 is an advanced α/β The counting system is designed to simplify the experimental operation and is suitable for long-time counting of large batch samples. One to four sample drawers are optional, and each sample drawer is configured with up to four 3.2cm or 5.7cm detectors.
It can collect water samples, biological samples, environmental samples and other samples α、 total β、 Activity measurement of Cs-137 and Pb-210. It is widely used in disease control and Prevention Center, water supply company, product quality supervision and inspection unit, environmental monitoring station, advanced laboratory and other units.
Canberra lb4200 is an advanced α/β The counting system is designed to simplify the experimental operation and is suitable for long-time counting of large batch samples. One to four sample drawers are optional, and each sample drawer is configured with up to four 3.2cm or 5.7cm detectors.
Less gas consumption and more excellent MDA value
With gas electronic monitoring system, it can automatically adjust the optimal gas pressure for the detector
Independent start / stop function of the detector provides maximum flexibility
Small footprint, easy to upgrade modular design
Automatic calibration: stable level, efficiency, self absorption and background
The detector can be used without removing the lead shield
The 7 cm diameter detector has the lowest background value in the industry
Each detector has an independent window to facilitate field replacement
Background:
Total count: guaranteed value: ≤ 0.8cpm
Typical value: < 0.7cpm
α Count: guaranteed value: ≤ 0.05cpm
Typical value: < 0.03cpm
β Count: guaranteed value: ≤ 0.75cpm
Typical value: < 0.67cpm
Efficiency: α (210Po):≥40%
β (90Sr-90Y):≥45%
String ratio: α reach β:< 1%
Detector stability level:
α (210Po)≤2.5%/100V:≥800V;
β (90Sr-90Y)≤2.5%/100V:≥200V