Alpha Analys™ Integrated Alpha Spectrometer

Alpha spectrometer is mainly used in water, biology, food, air and other samples α Nuclide analysis generally includes Am-241, Po-210, Pu-239 and other nuclides, which are widely used in environmental protection, nuclear power station, nuclear industry, health, customs, food and Drug Administration and other units.

Alpha analyst of Canberra uses apex alpha high efficiency laboratory software. Apex alpha is the most modern alpha spectroscopy software package, which can automatically complete a series of tasks such as alpha analyzer detector setting and calibration, quality assurance, sample analysis and reporting.

Share

Related Products

 
 

Alpha spectrometer is mainly used in water, biology, food, air and other samples α Nuclide analysis generally includes Am-241, Po-210, Pu-239 and other nuclides, which are widely used in environmental protection, nuclear power station, nuclear industry, health, customs, food and Drug Administration and other units.

Alpha analyst of Canberra uses apex alpha high efficiency laboratory software. Apex alpha is the most modern alpha spectroscopy software package, which can automatically complete a series of tasks such as alpha analyzer detector setting and calibration, quality assurance, sample analysis and reporting.

Provide fully integrated alpha spectroscopy instrument

Equipped with 100% computer controlled electronic and vacuum equipment

Realize automatic recoil suppression control

Modular structure with easy expansion

Provide a simple and clear graphical user interface

Clean exhaust ports shall be provided to reduce the accumulation of moisture and pollutants in the measuring room

Built in function to save a wide range of QA records

Able to directly connect to Ethernet network

GB / T 16141-95 determination of radionuclides α Energy spectrum analysis method

GB 14883.5-2016 national food safety standard Determination of radioactive substance polonium-210 in food

GB / T 12376-90 analytical method of polonium-210 in water - electroplating sample preparation method

GB / T 16141-95 determination of radionuclides α Energy spectrum analysis method

HJ 813-2016 analytical method for polonium-210 in water

GB 14883.8-2016 national standard for food safety determination of radioactive substances plutonium-239 and plutonium-240 in food
Method for radiochemical analysis of plutonium in water and soil samples (HJ 814-2016)

Dual alpha spectrometer module

  • Module - two vacuum chambers per spectrometer module;

  • Vacuum chamber - low background stainless steel;

  • Sample detector spacing - 1 to 45 mm in 4 mm increments

  • Maximum detector size - 1200 mm2
  • Vacuum range - 0 to > 26.7 kPa (0 to > 200 Torr) 
  • Detector leakage monitor – 4 Na to + 2500 Na; Read by the host computer.
  • Reference pulse - each counting chamber in the spectrometer module has its own independent pulse generator. Each pulse generator associated with all spectrometer modules is synchronized, allowing QA inspection of up to 48 counting cells at the same time.

Spectrometer performanc

  • Based on the use of a 450-18am pips detector and a high-quality 241Am point source
  • Energy resolution - 25%, detector source distance less than 1 mm
  • Background - above 3 MeV, < 1 count / hour
  • Frequency – 40.69 Hz. – Frequency stability – ± 50 ppm. Range – 0 ~ 10 MeV; Limited by the system threshold to 70% of the full range; Computer control. Adjust resolution – 12 bits (2.5 keV granularity).

Recoil suppression

 

  • Suppression bias voltage: + 5V is continuously applied to the detector shell to resist recoil contamination. 
  • Pressure range: 13-2.67 kPa (1-20 Torr), corresponding to air thickness of approximately 12 to 16 g / cm2 and detector sample distance of 5 mm to 45 mm

 

Spectrometer operation

  • System LLD - fixed at 5 V

     

  • MCA spectral length - 4096, 2048, 1024, 512, or 256

     

  • Preset mode - each spectrometer is preset independently; Actual or active time, multiple of 1s: maximum preset > 4x107s; Compound DT for all spectrometers in one cabinet (live time correction is the same for all spectrometers in one cabinet)

     

  • Control interface - Ethernet aim protocol, 2 / 802.3 compatible; Thin cable interface (10 B 2) built in; External Ethernet transceiver using AUI

     

  • MCA spectral length – 4096, 2048, 1024, 512, or 256.

     

  • MCA memory capacity – for 4096 channel ADC range, the capacity of each channel is 65535 (16 bits); For all other ranges, the capacity is 32 bits / track (> 4 x 109 counts / track)

 